Duo™ Service Manual
0070-10-0604-01
4 - 5
Repair Information
Troubleshooting Guide
Module Failures
MESSAGE/
PROBLEM REASON
SOLUTION
NIBP cuff cannot
be inflated.
Pinched or leaking hose
or cuff
Check hose and cuff. Replace as needed
Intermittently won't
take an NIBP
reading.
Loose cuff or patient
movement
Keep the patient quiet. Reapply cuff
NIBP readings
inappropriately
high or low for
patient condition.
Incorrect cuff size.
Incorrectly applied cuff
Use appropriate size cuff. Ensure correct cuff
application
NIBP module is out of
calibration
Calibrate/replace NIBP module
No SpO
2
reading
SpO
2
sensor or cable
damaged or
disconnected
Check sensor placement and connection. Replace
if damaged
Sensor not on patient
SpO
2
value is
inaccurate
SpO
2
sensor or cable
damaged or
disconnected
Check sensor placement and connection. Replace
if damaged
Sensor not on patient
Coloring agent (dye)
has been injected into
patient
Retry after the coloring agent has dissipated
Patient movement
Keep patient quiet
Patient is cold
Warm patient and retry
Patient is wearing nail
polish
Remove nail polish
Содержание datascope duo
Страница 1: ...Service Manual 0070 01 0604 02_revB_Duo srv color indd 1 3 10 10 5 17 19 PM...
Страница 2: ...Service Manual...
Страница 5: ...Table of Contents ii 0070 10 0604 01 Duo Service Manual This page intentionally left blank...
Страница 29: ...Introduction Parts 3 8 0070 10 0604 01 Duo Service Manual FIGURE 3 7 Battery Connector Assembly Detail...
Страница 33: ...Parts Listing Parts 3 12 0070 10 0604 01 Duo Service Manual This page intentionally left blank...
Страница 41: ...0070 10 0604 01 Revision F July 12 2010...