Calibration and System Checks
Calibration and Performance Verification
2 - 4
0070-10-0604-01
Duo™ Service Manual
2.4.4
Software Version Mode
Use the following procedure to view the software version.
1.
Ensure that the
Duo
is powered OFF.
2.
Press and hold the
Patient Size
key.
3.
While continuing to hold the
Patient Size
key, press and hold the
Power ON/OFF
key for two (2) seconds until the
Duo
beeps.
4.
Release both keys.
5.
After an additional 2-second delay,
Duo
will display “100” in the Pulse Rate tile and a
number in the Information Codes tile.
6.
Pressing the
Clear
key will cause the number displayed in the Pulse Rate tile to cycle
through a sequence of four numbers indicating which software version is being
displayed in the Information Codes tile as shown in the following table.
7.
To return to normal operation, press the
Power ON/OFF
key for two (2) seconds to
turn the
Duo
off.
8.
You may then turn the
Duo
back on to resume normal operation.
NOTE:
The Duo cannot be placed directly back into normal
monitoring mode from Software Version Mode. It must first
be powered OFF.
PULSE RATE TILE
INFORMATION CODES TILE SHOWS
100
Host Software Revision Level
200
NIBP Software Revision Level
300
SpO2 Software Revision Level
Содержание datascope duo
Страница 1: ...Service Manual 0070 01 0604 02_revB_Duo srv color indd 1 3 10 10 5 17 19 PM...
Страница 2: ...Service Manual...
Страница 5: ...Table of Contents ii 0070 10 0604 01 Duo Service Manual This page intentionally left blank...
Страница 29: ...Introduction Parts 3 8 0070 10 0604 01 Duo Service Manual FIGURE 3 7 Battery Connector Assembly Detail...
Страница 33: ...Parts Listing Parts 3 12 0070 10 0604 01 Duo Service Manual This page intentionally left blank...
Страница 41: ...0070 10 0604 01 Revision F July 12 2010...