SpO
2
Verification
Calibration and Performance Verification
2 - 8
0070-10-0604-01
Duo™ Service Manual
2.7
SpO
2
Verification
2.7.1
Test Equipment
• SpO
2
simulator
2.7.2
Test Procedure
1.
Connect the appropriate SpO
2
probe connector to the
Duo
monitor.
2.
Connect the SpO
2
probe to the SpO
2
simulator.
3.
Set the simulator target values to:
SpO
2
= 98%
Pulse Rate = 70
4.
Verify that the displayed SpO
2
and pulse rate values on the
Duo
monitor are +/- 2% of
the simulator target values.
5.
Change the simulator values.
6.
Verify the displayed values on the
Duo
monitor are equal to the simulator /- 2%.