background image

Parts Listing

Parts

3 - 10

0070-10-0604-01

Duo Service Manual

3.2

Parts Listing

REF. NUMBER

PART NUMBER

DESCRIPTION

1

0211-00-0146

Housing Screw (metric panhead)

2

0380-00-0475

Rear Housing

3

0441-00-0107

Chassis

4

0671-00-0045

Power Supply Board

5

0211-00-0145

Metric Panhead Screw

6

0671-00-0044

CPU/Display Board

7

0213-00-0032

Self Tapping Screw

8

0380-00-0472

Nellcor Connector Shroud

8A

0380-00-0473

Masimo Connector Shroud

9

0380-00-0476

Front Housing

10

0012-00-1595

SpO

2

 Power Cable

11

0671-00-0246

Masimo Isolated Power Board

12

0211-00-0143

Screw

13

0671-00-0243

Masimo SpO

2

 Board

14

0012-00-1474

Masimo Flex Cable

15

0671-00-0247

Nellcor Isolated Power Board

16

0671-00-0242

Nellcor SpO

2

 Board

16a

0671-00-0066

Nell-3 SpO

2

 Board

17

0012-00-1457

Nellcor Flex Cable

17a

0012-00-1661

Nellcor Flex Cable

18

0386-00-0308

SpO

2

 Mounting Plate

19

0104-00-0037

NIBP Module

20

0671-00-0063

Fan Driver Board

21

0671-00-0043

Battery Connector Board

22

0380-00-0481

Battery Latch

23

0380-00-0474

Battery Door

24

0346-00-0052

Battery Door Tether

25

0012-00-1592

Fan with cable

26

0348-00-0216

Fan Gasket

27

0213-00-4014

Screw

28

0386-00-0310

Fan Mounting Plate

29

0380-00-0471

Filler Panel

30

0367-00-0084

Handle

31

0211-00-0147

Handle Screw (metric panhead)

32

0334-00-1603-03

Rear Label, Lower, S/N

33

0348-00-0202

Foot

34

0380-00-0480-01

Power Switch Plunger

35

0380-00-0480-02

Clear Switch Plunger

36

0380-00-0480-03

Patient Size Switch Plunger

37

0380-00-0480-04

Start NIBP Switch Plunger

N/S - Not Shown

Содержание datascope duo

Страница 1: ...Service Manual 0070 01 0604 02_revB_Duo srv color indd 1 3 10 10 5 17 19 PM...

Страница 2: ...Service Manual...

Страница 3: ...igator is a U S trademark of Mindray DS USA Inc Masimo SET LNOP and CleanShield are U S registered trademarks of Masimo Corp Copyright Mindray DS USA Inc 2008 All rights reserved Contents of this publ...

Страница 4: ...tem Checks 2 3 Device Appearance and Installation Checks 2 3 Maintenance Functions Non Monitoring Modes 2 3 Unit of Measure Mode 2 3 Software Version Mode 2 4 Safety Test 2 5 Test Equipment 2 5 Case L...

Страница 5: ...Table of Contents ii 0070 10 0604 01 Duo Service Manual This page intentionally left blank...

Страница 6: ...iate areas throughout this manual A WARNING is provided to alert the user to potential serious outcomes death injury or serious adverse events to the patient or the user A CAUTION is provided to alert...

Страница 7: ...Introduction Warnings Precautions And Notes iv 0070 10 0604 01 Duo Service Manual This page intentionally left blank...

Страница 8: ...ompact lightweight portable patient monitor intended for monitoring the following vital signs blood pressure SpO2 optional and pulse rate on adult and pediatric patients The Duo monitor can be powered...

Страница 9: ...battery charging circuit will actively charge the battery while the Duo is connected to an AC source Battery charging takes place whether the monitor is on off or in use The power supply will automat...

Страница 10: ...e across C5 Primary Rectified Voltage DC Range 105 374 volts C5 Negative Lead Primary Ground Measure across Q1 pin1 and C5 Negative Lead 110k Hz Drive frequency Measure across C12 10 5 vdc input for F...

Страница 11: ...er commands from the Keypad The power supply board provides 3 3 vdc and 12 vdc to the CPU board These voltages are monitored by an A D converter located on the CPU board The CPU also controls an integ...

Страница 12: ...onnecting hose to the pressure transducer in the NIBP module The output of the pressure transducer is an analog pulsating signal This signal is filtered by a high pass filter and then amplified The am...

Страница 13: ...ne red and one infrared and a photodetector The two LEDs are alternately lighted at a precise frequency When the capillary vessels of the fingertip are filled a certain amount of light from the LEDs i...

Страница 14: ...roduction The following procedures are provided to verify the proper operation of the Duo monitor A menu driven interface is used to execute all verification tests Performance tests should be performe...

Страница 15: ...they are given 3 Use extreme care when reaching inside the opened instrument Do not contact exposed metal parts that may become live 4 Read through each step in the procedure so it is understood prio...

Страница 16: ...if used 2 4 2 Maintenance Functions Non Monitoring Modes When entering the maintenance functions non monitoring mode the monitor will perform a self test however the verification of functional LEDs wi...

Страница 17: ...er in the Information Codes tile 6 Pressing the Clear key will cause the number displayed in the Pulse Rate tile to cycle through a sequence of four numbers indicating which software version is being...

Страница 18: ...ground lead of the analyzer to the equipotential lug of the Duo monitor 3 Perform the leakage tests under the following conditions a Case grounded Normal polarity Normal polarity with open neutral Rev...

Страница 19: ...500 cc test chamber to the NIBP fitting on the Duo monitor under test 2 From the NIBP Calibration Mode code 525 momentarily press the CLEAR button on the Duo keypad The Duo will then switch to the Pn...

Страница 20: ...ual measured values displayed on the Duo monitor may not compare with the selected target pressure on the simulator This test is intended to confirm the REPEATABILITY not accuracy of dynamic NIBP read...

Страница 21: ...the appropriate SpO2 probe connector to the Duo monitor 2 Connect the SpO2 probe to the SpO2 simulator 3 Set the simulator target values to SpO2 98 Pulse Rate 70 4 Verify that the displayed SpO2 and p...

Страница 22: ...Duo Service Manual 0070 10 0604 01 3 1 3 0 Parts 3 1 Introduction This section contains exploded views of the Duo monitor internal modules and parts list...

Страница 23: ...Introduction Parts 3 2 0070 10 0604 01 Duo Service Manual FIGURE 3 1 Duo Exploded View SpO2 Module See SpO2 Detail 8A 7 9 8 5 3 2 1 4 6 5...

Страница 24: ...Duo Service Manual 0070 10 0604 01 3 3 Parts Introduction FIGURE 3 2 Masimo SpO2 Detail 10 11 12 13 12 14 19 10 11 12 13 12 18 14...

Страница 25: ...Introduction Parts 3 4 0070 10 0604 01 Duo Service Manual FIGURE 3 3 Nellcor SpO2 Detail 16 17 18 12 10 15 16a 17a...

Страница 26: ...Duo Service Manual 0070 10 0604 01 3 5 Parts Introduction FIGURE 3 4 Main Frame 5 Battery Latch Detail 4 5 3 5 20 21 19 22 11...

Страница 27: ...Introduction Parts 3 6 0070 10 0604 01 Duo Service Manual FIGURE 3 5 Rear Case Assembly 26 27 28 29 30 31 1 32 33 34 35 36 45 25 26 27 28 29 24 33 23 35 41 31 30 1...

Страница 28: ...Duo Service Manual 0070 10 0604 01 3 7 Parts Introduction FIGURE 3 6 Front Case Assembly 9 37 38 39 40 41 42 43 9 35 34 36 37 38 40 39...

Страница 29: ...Introduction Parts 3 8 0070 10 0604 01 Duo Service Manual FIGURE 3 7 Battery Connector Assembly Detail...

Страница 30: ...LSTD Duo monitor mounting kit 0406 00 0857 01 Casters Non locking 0401 00 0045 Casters Locking 0401 00 0046 Utility basket 0202 00 0166 Included in Duo monitor mounting kit Casters locking Mounting bo...

Страница 31: ...lated Power Board 16 0671 00 0242 Nellcor SpO2 Board 16a 0671 00 0066 Nell 3 SpO2 Board 17 0012 00 1457 Nellcor Flex Cable 17a 0012 00 1661 Nellcor Flex Cable 18 0386 00 0308 SpO2 Mounting Plate 19 01...

Страница 32: ...e below Display Overlay N S See table below Connector Label Display Overlay LANGUAGE OPTION PART NUMBER English NIBP Only 0330 00 0053 01 English NIBP SpO2 0330 00 0053 11 Upper Rear Label LANGUAGE PA...

Страница 33: ...Parts Listing Parts 3 12 0070 10 0604 01 Duo Service Manual This page intentionally left blank...

Страница 34: ...nderstanding of the instrument functions as well as an understanding of the theory of operation FIGURE 4 1 Module Interconnection P2 P1 Fan Drive Board Fan Battery P1 Power Supply Board P2 P3 Serial I...

Страница 35: ...ded the specified upper safety limit E12 NIBP System Failure Operation of blood pressure pump system has failed E13 NIBP Time Out Measuring time has exceeded 120 seconds E14 NIBP Illegally Reset Unexp...

Страница 36: ...ng problems communicating i e framing errors or bad checksums with the Nellcor module E43 Nellcor SpO2 Weak Pulse A pulse rate can not be determined and all other measurement conditions are normal The...

Страница 37: ...PROBLEM REASON SOLUTION No display after power on power indicator does not light Bad line fuse Replace fuse Bad power supply Replace power supply Bad CPU Display board Replace CPU Display board NIBP...

Страница 38: ...d cuff Use appropriate size cuff Ensure correct cuff application NIBP module is out of calibration Calibrate replace NIBP module No SpO2 reading SpO2 sensor or cable damaged or disconnected Check sens...

Страница 39: ...the NIBP tubing from the front housing pressure fitting 4 3 3 SpO2 Interface Board Removal 1 Disconnect the cable from P01 on the CPU Display pcb 2 Remove three 3 5mm hexnuts from their stand offs on...

Страница 40: ...head machine screws from the corners of the power supply PCB 3 Remove one 1 3 x 6 mm Phillips panhead machine screw from the line power entry connector 4 Slide the Power Supply pcb towards the bottom...

Страница 41: ...0070 10 0604 01 Revision F July 12 2010...

Страница 42: ...621 Mindray UK Limited 3 Percy Road St John s Park Huntingdon Cambridgeshire PE29 6SZ United Kingdom Tel 01480 416840 Fax 01480 436588 Mindray Medical France SARL Europarc Cr teil 123 Chemin des Bass...

Отзывы: