Duo™ Service Manual
0070-10-0604-01
1 - 3
Theory of Operation
Hardware Overview
FIGURE 1-2
Power Supply Block Diagram
Power Supply Voltage Test Points:
Location
Function
Measure across C5
Primary Rectified Voltage (DC). Range: 105 - 374 volts.
C5 Negative Lead
Primary Ground.
Measure across Q1 pin1 and
C5 Negative Lead
110k Hz Drive frequency.
Measure across C12
+10.5 vdc input for Fan Driver board.
C12 Negative Lead
Secondary Ground.
C47 Positive Lead
+5 vdc supply for Power On/Off Control.
C50 Positive Lead
+3.3 vdc output.
C68 Positive Lead
+12 vdc output.