Troubleshooting Guide
Repair Information
4 - 2
0070-10-0604-01
Duo™ Service Manual
4.2
Troubleshooting Guide
Error Codes and Solutions
MESSAGE/
PROBLEM
REASON
SOLUTION
E01
NIBP Self Test Error
NIBP Module hardware failure
E02
NIBP Communications
Error
Communications with NIBP Module have failed
E03
Loose Cuff
Cuff is not properly wrapped or no cuff is present
E04
Air Leak
Cuff, hose or connector is damaged, internal leak
E05
Air Pressure Failure
Stable pressure value is not available (e.g. hoses
are pinched or occluded)
E06
Successful Pneumatic
Test
Indicates NIBP pneumatic test was successful
E07
Pneumatic Test Failed/
Pneumatic Leak
Leak detected during the pneumatic test
E08
Weak Signal
Cuff is too loose or patient pulse is too weak
E09
Range Exceeded
NIBP value exceeds the upper measurement limit
E10
Excessive Motion
Signal Saturated
Monitor is detecting too much motion and/or
noise to obtain a reading
E11
Over Pressure
Pressure has exceeded the specified upper safety
limit
E12
NIBP System Failure
Operation of blood pressure pump system has
failed
E13
NIBP Time Out
Measuring time has exceeded 120 seconds
E14
NIBP Illegally Reset
Unexpected NIBP reset
E15
NIBP Reset Failed
NIBP reset failed
E16
NIBP Communications
CRC Error
NIBP Serial Communication CRC failure
E17
NIBP Patient Size
Change Error
Attempt to change Patient Size failed
E20
Masimo SpO
2
Interference
Noise detected on the pulse signal prevents pulse
discrimination
E21
Masimo SpO
2
Low
Perfusion
Patient perfusion is low
E22
Masimo SpO
2
Too
Much Light
There is too much ambient room light for the
sensor to function properly
E23
Masimo SpO
2
Unrecognized Sensor
The monitor does not recognize the sensor
E24
Masimo SpO
2
Communication Error
The monitor and the SpO
2
module are not
communicating
E25
Masimo SpO
2
Board
Fault
The Masimo SET board has failed to operate
properly
E26
Masimo SpO
2
Sensor
Fault
Defective sensor
E28
Masimo SpO
2
Timeout
SpO
2
data has been determined continuously for
more than 2 minutes, so SpO
2
data has timed out
from the display
E29
Masimo SpO
2
Low
Signal IQ
The SpO
2
signal quality is poor
Содержание datascope duo
Страница 1: ...Service Manual 0070 01 0604 02_revB_Duo srv color indd 1 3 10 10 5 17 19 PM...
Страница 2: ...Service Manual...
Страница 5: ...Table of Contents ii 0070 10 0604 01 Duo Service Manual This page intentionally left blank...
Страница 29: ...Introduction Parts 3 8 0070 10 0604 01 Duo Service Manual FIGURE 3 7 Battery Connector Assembly Detail...
Страница 33: ...Parts Listing Parts 3 12 0070 10 0604 01 Duo Service Manual This page intentionally left blank...
Страница 41: ...0070 10 0604 01 Revision F July 12 2010...