C-3
Alarm
Information
Alarm
Level
Reason
Countermeasure
[Standby Time
Expired]
Medium
Pump is in standby mode
and standby time is
completed.
Press
or
to
cancel alarm.
[Reminder]
Low
The infusion pump performs
no operation during the set
reminder time after the
infusion set is loaded.
1. Operate the pump
(except press
) to
cancel alarm.
2. Or press
to
acknowledge the alarm
condition.
[Low Battery]
Low
Only powered with built-in
battery, battery charge is
insufficient.
1. Connect to the AC
power source to cancel
alarm.
2. Or press
to
acknowledge the alarm
condition.
[Time Near End]
Low
Infusion remaining time
reaches the setting value of
[Time Near End].
1. The alarm will not be
cancelled automatically
until the infusion is
completed, and then
switch to [VTBI Complete]
or [Empty] alarm.
2. Or press
to cancel
alarm.
3. Or press
to
acknowledge the alarm
condition.
[No AC Power]
Low
Power cord disconnects
when the pump has
connected to an AC/DC
power supply.
1. Connect to an AC/DC
power supply to cancel the
alarm.
2. Or press
to cancel
alarm.
[Communication
interrupted]
Low
Infusion pump and
BeneFusion CS5 Infusion
Supervision System are
communicated successfully
Press
or restore the
communication between
infusion pump(s) and
Содержание BeneFusion VP3 Vet
Страница 1: ...BeneFusion VP3 Vet Infusion Pump Operator s Manual...
Страница 2: ......
Страница 20: ...2 4 2 2 2 Front View with the Door Opened 1 Liquid check clip button 2 Liquid check clip 3 Infusion set slot...
Страница 73: ...A 11 Unit of Rate ml h...
Страница 90: ...F 1 F Declaration of Conformity P N 046 013101 00 4 0...
Страница 91: ......
Страница 92: ......