C-1
C
Alarm Information
This chapter presents the alarm information of the infusion pump. Prompt
information for operation guidance will not be presented in this chapter.
The table shows the appropriate countermeasures for each piece of information
related to alarm triggering. If the problem still exists after operating according to the
countermeasures, please contact the company.
Alarm
Information
Alarm
Level
Reason
Countermeasure
[Occlusion]
High
Infusion cannula blocked
during infusion, and
occlusion pressure reaches
the threshold of preset
occlusion pressure threshold.
Infusion is stopped after the
alarm is triggered.
Press
to cancel
alarm, check and
eliminate the source of the
alarm.
[Air in line]
High
Size of single air detection
reaches to the preset value.
Infusion is stopped after the
alarm is triggered.
Press
to cancel
alarm, check and
eliminate the air in line.
[Accumulate air]
High
Size of accumulate air
bubble accumulated in 1
hour reaches to the preset
value. Infusion is stopped
after the alarm is triggered.
Press
to cancel
alarm, check and
eliminate the air in line.
[Drop error]
High
If drop sensor is installed
correctly and the switch of
[Drop sensor] is on, drop
sensor detects that the drop
rate deviates the preset
value. Infusion is stopped
after the alarm is triggered.
Press
to cancel
alarm, refill the drop
chamber correctly or place
the drop chamber
vertically or avoid the
strong light.
[No Battery]
High
1. Only powered with built-in
battery, battery is empty.
2. Built-in battery and
external DC power encounter
failure.
Connect to the AC power
source to cancel alarm.
Содержание BeneFusion VP3 Vet
Страница 1: ...BeneFusion VP3 Vet Infusion Pump Operator s Manual...
Страница 2: ......
Страница 20: ...2 4 2 2 2 Front View with the Door Opened 1 Liquid check clip button 2 Liquid check clip 3 Infusion set slot...
Страница 73: ...A 11 Unit of Rate ml h...
Страница 90: ...F 1 F Declaration of Conformity P N 046 013101 00 4 0...
Страница 91: ......
Страница 92: ......