7-1
7
Other Functions
7.1 History Record
The infusion pump when in use will produce some key data stored in [
History
Record
], providing foundation for the treatment review and maintenance review at a
later period. The attribute of recording events includes action, time and description.
A record is created once an event occurs. The memory can store up to 1500 records.
Once the memory is full, the oldest records will be removed first. History record will
not loss when the infusion pump powers down.
1. Select [
Main Menu
]→[
History Record
].
2. Select [
History Record
]: Each page can demonstrate up to 2 records, and
press
to turn the pages.
7.2 Power-down Save
To prevent the loss of animal data and alarm settings when the pump suddenly
powers down, the pump provides the function of the power-down data and alarm
settings storage. If the pump powers down suddenly after it is restarted, the last
infusion parameters will display the alarm information and will remain in consistency
with those before the power-down, and will be reloaded. You can refer to [
History
Record
] to view such information as infusion parameters and alarm.
7.3 Nurse Call
Select [
Main Menu
]→[
System Options
]→[
Nurse call
] , and set in the open menu:
Switch
On: Indicates switching on the nurse call function.
Off: Indicates switching off the nurse call function.
Signal type
Continuous: Indicates that the output nurse call signal type is the same as that of the
alarm existence time, i.e., from the occurrence of the alarm to the end of it.
Pulse: Indicates the output nurse call signal is a pulse signal with the type of 1
second. When several alarms exist at the same time, only one pulse signal can be
outputted. If the current alarm is not removed and another alarm occurs, then one
additional pulse signal is outputted.
Trigger type
Normally Close: Select when the hospital call system is set as [
NORM. Close
].
Normally Open: Select when the hospital call system is set as [
NORM. Open
].
Содержание BeneFusion VP3 Vet
Страница 1: ...BeneFusion VP3 Vet Infusion Pump Operator s Manual...
Страница 2: ......
Страница 20: ...2 4 2 2 2 Front View with the Door Opened 1 Liquid check clip button 2 Liquid check clip 3 Infusion set slot...
Страница 73: ...A 11 Unit of Rate ml h...
Страница 90: ...F 1 F Declaration of Conformity P N 046 013101 00 4 0...
Страница 91: ......
Страница 92: ......