4.2.11 Infusion Pause
During infusion, if changing the drug liquid or infusion set is needed, press
to
enter the [
Pause
] interface to stop the infusion. On the [
Pause
] screen, press
to return to the parameters setting interface, and you can modify infusion
parameters; Press
to continue the infusion.
4.2.12 BOLUS
In any run screen in the infusion mode, press
to enter the [
Bolus
] settings
screen. There are two ways to start the bolus:
Manual Bolus: Set [
Bolus Rate
],
press and hold
to manual bolus, and
release to return to the original rate.
Auto Bolus: Set [
Bolus Rate
], [
Bolus VTBI
] and [
Bolus Time
], press
to
auto bolus.
NOTE
If no operation is performed within 2 minutes, the infusion pump will
automatically exit the Bolus Settings screen and the procedure must be
repeated.
Animal's clinical condition and working condition of the infusion pump
must be monitored carefully.
4.2.13 Change the Rate during Operation
Rate may be modified without stopping the infusion. In any running screen of the
infusion mode, press
to change the value of the [
Rate
] into the adjustable
state, thus to set the expected rate, press
or
again for confirmation,
then start to infuse under the new set rate.
Содержание BeneFusion VP3 Vet
Страница 1: ...BeneFusion VP3 Vet Infusion Pump Operator s Manual...
Страница 2: ......
Страница 20: ...2 4 2 2 2 Front View with the Door Opened 1 Liquid check clip button 2 Liquid check clip 3 Infusion set slot...
Страница 73: ...A 11 Unit of Rate ml h...
Страница 90: ...F 1 F Declaration of Conformity P N 046 013101 00 4 0...
Страница 91: ......
Страница 92: ......