A-12
Unit of Rate (gtt/min)
Sampling rate: 20gtt/min
Sampling interval:
△
t =1min
Test period: t =120mins
Infusion rate: Q (gtt/min)
Sampling rate: 20gtt/min
Sampling interval:
△
t =1min
Observation windows: p
△
t =2, 5,
11, 19, 31mins
Maximum deviation over the
course of a full observation
window: Ep(Max) (%)
Minimum deviation over the
course of a full observation
window: Ep(Min) (%)
Average deviation: A (%)
Sampling rate: 20gtt/min
Sampling interval:
△
t =1min
Observation windows: p
△
t =2, 5,
11, 19, 31mins
Maximum deviation over the
course of a full observation
window: Ep(Max) (%)
Minimum deviation over the
course of a full observation
window: Ep(Min) (%)
Average deviation: B(%)
Содержание BeneFusion VP3 Vet
Страница 1: ...BeneFusion VP3 Vet Infusion Pump Operator s Manual...
Страница 2: ......
Страница 20: ...2 4 2 2 2 Front View with the Door Opened 1 Liquid check clip button 2 Liquid check clip 3 Infusion set slot...
Страница 73: ...A 11 Unit of Rate ml h...
Страница 90: ...F 1 F Declaration of Conformity P N 046 013101 00 4 0...
Страница 91: ......
Страница 92: ......