B - 3
NOTE
•
The device needs special precautions regarding EMC and needs to be
installed and put into service according to the EMC information provided
below.
•
Use of portable or mobile communications devices will degrade the
performance of the device.
•
Other devices may affect this device even though they meet the
requirements of CISPR.
•
If the essential performance is lost or degraded, it may be necessary to take
mitigation measures, such as re-orienting or relocating the ME EQUIPMENT
or ME SYSTEM or shielding the location or stopping using the pump system
and contact the service personnel.
If the device is operated within the electromagnetic environment listed in Table
Guidance and Declaration —Electromagnetic Immunity
, the system will remain safe
and provide the following essential performance:
■
Operating mode
■
Accuracy
■
Function
■
Protection against UNINTENDED BOLUS volumes
■
Occlusion
■
ALARM CONDITIONS regarded
■
Data stored
Содержание BeneFusion uVP
Страница 1: ...BeneFusion uVP BeneFusion uVP ex Infusion Pump Instruction for Use ...
Страница 2: ......
Страница 12: ...1 4 This page intentionally left blank ...
Страница 18: ...3 4 This page intentionally left blank ...
Страница 26: ...6 2 This page intentionally left blank ...
Страница 30: ...8 2 This page intentionally left blank ...
Страница 32: ...9 2 This page intentionally left blank ...
Страница 36: ...10 4 This page intentionally left blank ...
Страница 38: ...11 2 This page intentionally left blank ...
Страница 40: ...12 2 This page intentionally left blank ...
Страница 50: ...A 10 A 7 Operating Environment Operating system FreeRTOS Classification OS Core Version Information 9 0 0 ...
Страница 60: ...B 10 This page intentionally left blank ...
Страница 61: ......
Страница 62: ...KF H 046 024481 00 4 0 ...