Do not open the equipment housings. All servicing and future upgrades must
be carried out by trained and authorized personnel. Moreover, the servicing
must be done only after the AC power supply is disconnected.
•
Do not place the equipment or accessories in any position that might cause it
to fall on the patient.
•
Do not start an infusion unless the setup was verified to be correct.
•
To avoid inadvertent disconnection, route all cables in a way to prevent a
stumbling hazard. Wrap and secure excess cabling to reduce risk of
entanglement by patients or personnel.
•
Clearing the occlusion result from line kinks, filter coagulation, etc. may
cause extra bolus to patients. Appropriate measures should be taken.
•
Do not touch the patient and device connectors simultaneously. Otherwise
leakage current may result in patient injury.
•
To avoid electric shock, do not touch patient and other non-defibrillation
proof equipments during defibrillation. Defibrillation will not affect the
performance of the equipment.
1.1.2
Cautions
CAUTION
•
When several infusion lines are connected to the same vascular access, there
may be back flow or prolonged response time of occlusion alarm. Therefore,
use check valve at the line end or follow local hospitals’ instructions while in
connection with other infusion system.
•
When using this equipment for enteral nutrition, do not use enteral fluids for
intravenous infusion to avoid patient injury, and use only dedicated
disposable enteral feeding sets for enteral nutrition.
•
Ensure that the equipment is supplied with continuous electric power during
work. Sudden power failure may cause data loss.
•
Electromagnetic fields may affect equipment performance. This makes it
necessary for other equipment used in the vicinity of this equipment to meet
EMC standards. Mobile phones, X ray and MRI equipment are all potential
interference sources because of their high-intensity electromagnetic
radiation.
•
Always install or carry the equipment properly to avoid damage caused by
drop, impact, strong vibration or other mechanical force. The equipment
should be observed to verify normal operation after fall, otherwise it cannot
be used.
•
Dry the equipment immediately in case of rain or water spray.
Содержание BeneFusion uVP
Страница 1: ...BeneFusion uVP BeneFusion uVP ex Infusion Pump Instruction for Use ...
Страница 2: ......
Страница 12: ...1 4 This page intentionally left blank ...
Страница 18: ...3 4 This page intentionally left blank ...
Страница 26: ...6 2 This page intentionally left blank ...
Страница 30: ...8 2 This page intentionally left blank ...
Страница 32: ...9 2 This page intentionally left blank ...
Страница 36: ...10 4 This page intentionally left blank ...
Страница 38: ...11 2 This page intentionally left blank ...
Страница 40: ...12 2 This page intentionally left blank ...
Страница 50: ...A 10 A 7 Operating Environment Operating system FreeRTOS Classification OS Core Version Information 9 0 0 ...
Страница 60: ...B 10 This page intentionally left blank ...
Страница 61: ......
Страница 62: ...KF H 046 024481 00 4 0 ...