4 - 1
4
Getting Started
4.1
Turning on the Pump
WARNING
•
Before putting the system into operation, the operator must verify that the
equipment, connecting cables and accessories are in correct working order
and operating condition.
•
Check that visual and auditory alarm signals are presented correctly when
the equipment is powered on. Do not use the equipment if you suspect it is
not working properly, or if it is mechanically damaged. Contact your service
personnel or us.
NOTE
•
Stay within 1 meter (39 inches)
of the pump while setting it up and operating
it, making sure that you have a clear view of the pump interface.
•
The equipment uses a mains plug as isolation means to the mains power. Do
not locate the equipment in a place difficult to operate the mains plug.
4.2
Loading the Infusion Set
WARNING
•
To ensure the accuracy of air bubble detection, check and remove the
remained fluid in the infusion set slot before loading the infusion set.
•
While loading the infusion set, do not touch the free-flow clamp to avoid
being hurt.
•
This pump uses standard, single use infusion set with Luer lock connections.
•
The pump must be mounted to the same level as the patient’s heart. The most
accurate pressure monitoring in the infusion set is achieved when the pump
is positioned close to the patients heart level.
•
We recommend you to use an infusion set stated in this manual. If a non-
recommended infusion set must be used or a different set needs to be
changed, perform the calibration and performance test before use.
Содержание BeneFusion uVP
Страница 1: ...BeneFusion uVP BeneFusion uVP ex Infusion Pump Instruction for Use ...
Страница 2: ......
Страница 12: ...1 4 This page intentionally left blank ...
Страница 18: ...3 4 This page intentionally left blank ...
Страница 26: ...6 2 This page intentionally left blank ...
Страница 30: ...8 2 This page intentionally left blank ...
Страница 32: ...9 2 This page intentionally left blank ...
Страница 36: ...10 4 This page intentionally left blank ...
Страница 38: ...11 2 This page intentionally left blank ...
Страница 40: ...12 2 This page intentionally left blank ...
Страница 50: ...A 10 A 7 Operating Environment Operating system FreeRTOS Classification OS Core Version Information 9 0 0 ...
Страница 60: ...B 10 This page intentionally left blank ...
Страница 61: ......
Страница 62: ...KF H 046 024481 00 4 0 ...