B - 5
Guidance and Declaration - Electromagnetic Immunity
The device is intended for use in the specified electromagnetic environment. The customer
or the user of the device should assure that it is used in such an environment as described
below.
Immunity
test
IEC 60601
Test level
Compliance
level
Electromagnetic environment -
guidance
Conducted
disturbances
induced by RF
fields
IEC61000-4-6
3 Vrms
150 kHz to
80 MHz
3 Vrms
Portable and mobile RF communications
equipment should be used no closer to
any part of the device, including cables,
than the recommended separation
distance calculated from the equation
applicable to the frequency of the
transmitter. Recommended separation
distance:
150kHz to 80 MHz
150kHz to 80 MHz
80 MHz to 800 MHz
800 MHz to 2.7 GHz
where P is the maximum output power
rating of the transmitter in watts (W)
according to the transmitter
manufacturer and d is the recommended
separation distance in meters (m).
Field strengths from fixed RF
transmitters, as determined by an
electromagnetic site survey
b
, should be
less than the compliance level in each
frequency range
c
.
Interference may occur in the vicinity of
equipment marked with the following
symbol:
6 Vrms
in ISM
bandsa
between
0,15 MHz
and 80 MHz
a
6 Vrms
Radiated RF
EM fields
IEC61000-4-3
10V/m
80 MHz to
2.7 GHz
10V/m
Note 1: At 80 MHz and 800 MHz, the higher frequency range applies.
Note 2: These guidelines may not apply in all situations. Electromagnetic propagation is
affected by absorption and reflection from structures, objects and people.
P
d
2
.
1
=
P
d
2
.
1
=
P
d
3
.
2
=
Содержание BeneFusion uVP
Страница 1: ...BeneFusion uVP BeneFusion uVP ex Infusion Pump Instruction for Use ...
Страница 2: ......
Страница 12: ...1 4 This page intentionally left blank ...
Страница 18: ...3 4 This page intentionally left blank ...
Страница 26: ...6 2 This page intentionally left blank ...
Страница 30: ...8 2 This page intentionally left blank ...
Страница 32: ...9 2 This page intentionally left blank ...
Страница 36: ...10 4 This page intentionally left blank ...
Страница 38: ...11 2 This page intentionally left blank ...
Страница 40: ...12 2 This page intentionally left blank ...
Страница 50: ...A 10 A 7 Operating Environment Operating system FreeRTOS Classification OS Core Version Information 9 0 0 ...