22
●
[
F-07
]
Flow reference standard selection
Select the flow value is shown under standard or normal
condition.
< Operation >
Press or button at function selection mode to display
[
F-07
] [
rEFE
].
●
[
F-08
]
Analog output selection
Select the analog output signal is for flow sensor or pressure
sensor.
< Operation >
Press or button at function selection mode to display
[
F-08
] [
AnG
].
Flow
reference
Analog
Selected
condition
Selected
sensor
Flow Reference Standard Selection
Sensor Selection
Press or button to select the
standard or normal condition.
Press or button to select the
sensor of analog output.
Standard conditon
(ANR)
Flow sensor
Normal condition
(NOR)
Pressure sensor
NOTE
1. Stardard condition (ANR)
The display value is calculated under 20°C, 1atm.
2. Normal condition (NOR)
The display value is calculated under 0°C, 1atm.
3. Flow rate in the specification is the value at standard condition (ANR).
NOTE
This function is not available with Output Specification -02 and -04.
Press
button
Press
button
FLOW & PRESSURE SENSOR
( INSTRUCTION MANUAL )
MFP01
How to use
–
Operation setting instructions
rEFE Anr
L/min
kPa
Ag
EEPr oFF
L/min
kPa
Ag
EEPr 2Aun
L/min
kPa
Ag
EEPr oFF
L/min
kPa
Ag
EEPr 5Aun
L/min
kPa
Ag
AnG FLoY
L/min
kPa
Ag
AnG PrES
L/min
kPa
Ag
AnG FLoY
L/min
kPa
Ag
rEFE nor
L/min
kPa
Ag
rEFE Anr
L/min
kPa
Ag
F-07 rEFE
L/min
kPa
Ag
F-09 EEPr
L/min
kPa
Ag
F-08 AnG
L/min
kPa
Ag
Press
button to return to
Function Selection Mode
Press
button to return to
Function Selection Mode
Press
button to return to
Function Selection Mode
●
[
F-09
]
Accumulated value hold selection
The default setting is "OFF", the accumulated flow value is
zeroed when the power supply is turned off.
Select this function to keep accumulated flow value to be
stored in permanent memory and reload the recent saved
accumulated value after power supply turns on.
< Operation >
Press or button at function selection mode to display
[
F-09
] [
EEPr
].
Accumulated
value hold
Selected
function
Accumulated Value Hold Selection
Press or button to turn off the
function or select the data stored cycle.
Function "OFF"
Data stored
every 2 min.
Data stored
every 5 min.
NOTE
The maximum writable limit of the memory device is 1 million cycles.
If the sensor is operated 24 hours per day, the durability is calculated
as below:
● 5 minutes x 1 million cycles = 5 million minutes = 9.5 years
● 2 minutes x 1 million cycles = 2 million minutes = 3.8 years
Press
button