Installation instructions
●
Piping
Install the pipe by following the arrow indication that shows
the air flow direction on the product.
Use straight piping 8cm or longer to connect the Piping Port
(Inlet side).
If straight piping is not installed, the accuracy may vary by ±
2% F.S..
* Straight Piping: The pipe is without bending and the cross
sectional areas of the pipe keeps the same.
Blow the air to flush out the foreign matters, dust and etc.
before installing the pipe.
Uncleaned air may cause malfunction or damage to the
product.
Piping for the One-Touch Fitting, insert the tube firmly into the
fitting and make sure it cannot be pulled out.
Also using the proper tube cutter is recommended to ensure
square edge tube.
●
Mounting Bracket / Optional Parts
The LCD display may be difficult to see at certain angles.
The sensor can be installed horizontally or vertically, but
the flow rates
may change because of the installation way of the product
or piping.
The tightening torque for screws should be under 0.5 ±0.1
N.m.
* The tightening torque for bracket mounting screws should
be under 0.5 ± 0.1 N.m.
Flow direction
straight piping
8cm or longer
Piping Port
(Inlet side)
Piping Port
(Outlet side)
Horizontal mounting (by Through-Hole)
Bracket mounting
Mounting screw
Mounting screw
Mounting screw
P type
3×6L