APPENDIX A
ERROR CODES
1.
Numeric underflow: probably a bad U parameter or incorrect flow program P2.
2.
Numeric overflow: probably a bad U parameter or incorrect flow program P2. P9 to small, relay is
pulsing once every 2 seconds until the totalizer is overflowed at high flowrates.
3.
Attempt to divide by zero: check parameter entries.
4.
Bad argument (e.g. attempted square root of a negative number): check parameter entries.
5.
Attempt to access non-existent parameter.
6.
Access to parameters denied: security code has not been entered.
7.
Primary element exceeds BS-3680 dimensions or head exceeds the max. range of BS-3680
specifications; bad U parameter.
8.
Conversion overflow (long floating point to short floating point):
Inappropriate choice of flow or time units, (U0 or U1) causing logging total overflow.
Bad U parameter causing excessive flowrate to be calculated.
9.
Requested date is not in data log:
date has not been entered in correct format.
calendar was not set correctly on start up.
10.
Attempt to select non-existent primary element of P2
PL-269
9 – 1
Содержание OCM II PL-269
Страница 1: ...OCM II PROCESS MEASUREMENTS OPEN CHANNEL MONITOR Instruction Manual PL 269 February 1991 33452690 ...
Страница 6: ......
Страница 8: ......
Страница 12: ......
Страница 22: ......
Страница 28: ......
Страница 29: ......
Страница 30: ......
Страница 31: ......
Страница 32: ......
Страница 33: ......
Страница 34: ......
Страница 35: ......
Страница 36: ......
Страница 76: ......
Страница 78: ......
Страница 92: ...PL 269 9 14 ...
Страница 94: ......
Страница 95: ......
Страница 106: ......