TEMPERATURE SENSOR OUTLINE & WIRING
NOTES:
1. The temperature sensor should be mounted in a location which represents the temperature
fluctuations likely to occur between the transducer and the target.
2. To avoid false readings, mount the temperature sensor out of direct sunlight. Radiant heating can
cause a differential of 20 °C between air and sensor temperature.
3. Temperature sensor cable can be run with the transducer cable in a grounded metal conduit.
Ground shield only at OCM II. Insulate shield at junctions to prevent inadvertent grounding.
* 4. If a temperature sensor is used, the 9.53K
Ω
resistor must be removed from across terminals
TB-1 13 & 14 .
5. Temperature sensor wiring must be done in conjunction with approved conduit, boxes and fittings
and to procedures in accordance with all governing regulations.
FIG. 4
connection via customer’s junction box
OCM II *
13 14 15
SHLD
WHT
BLK
BLK
WHT
SHLD
BLK
WHT
SHLD
41 mm
(1.63")
1" NPT
cable
59 mm
(2.31")
PL-269
10 – 5
Содержание OCM II PL-269
Страница 1: ...OCM II PROCESS MEASUREMENTS OPEN CHANNEL MONITOR Instruction Manual PL 269 February 1991 33452690 ...
Страница 6: ......
Страница 8: ......
Страница 12: ......
Страница 22: ......
Страница 28: ......
Страница 29: ......
Страница 30: ......
Страница 31: ......
Страница 32: ......
Страница 33: ......
Страница 34: ......
Страница 35: ......
Страница 36: ......
Страница 76: ......
Страница 78: ......
Страница 92: ...PL 269 9 14 ...
Страница 94: ......
Страница 95: ......
Страница 106: ......