ST-25 TRANSDUCER OUTLINE AND WIRING
NOTES:
- Do not route cable openly in raceways
- Cable length:
- standard:
- 1m (3’)
- optional:
- consult Milltronics
- Radiating surface:
- polyurethane
- Cable must be run in a grounded metal conduit with no other cabling (except temp. sensor cable).
Ground shield only at OCM II. Insulate shield at junctions to prevent inadvertent grounding.
- Transducer wiring must be done in conjunction with approved conduit, boxes and fittings and to
procedures in accordance with all governing regulations.
ST -25 transducer
11 12
11 12
120 mm
(4.7")
1" NPT
black
white
hot
black
white
shield
connection via customer’s
junction box
OR
direct connection
radiating surface
84 mm
(3.3") dia.
TB 1
TB 1
OCM II
cable
shield
RG62U
coaxial cable
FIG. 3
PL-269
10 – 4