TRANSDUCER MOUNTING - OCM II APPLICATIONS
Bracket
flexible conduit
* 1" NPT PVC
coupling
steel channel
ST-25
transducer
Flexible conduit mounted transducer should not be
subjected to wind, vibration or jarring.
Flexible Conduit
* 1" NPT PVC nipple
Blind Flange
Plywood
* 1" NPT
PVC coupling
Plywood mounting provides
excellent isolation, but must be rigid
enough to avoid flexing if subjected
to loading.
Flange, gasket and hardware supplied by customer.
Refer to Figure 7.
Guidelines:
1. Do not mount transducer directly to metal. Use PVC coupling and nipple as provided.
2. Do not overtighten mounting. Hand tightening PVC coupling and nipple is sufficient.
3. In OCM II applications, transducer mount (flange, bracket, conduit or plywood) must keep
transducer perpendicular to liquid surface.
4. Transducer installation, whether in non-hazardous (as shown) or hazardous areas, must be done
in conjunction with approved conduit, boxes and fittings and to procedures in accordance with all
governing regulations.
5. Inner standpipe surface and ends must be smooth and free of burrs, ridges or seams.
FIG. 5
PL-269
10 – 6
Содержание OCM II PL-269
Страница 1: ...OCM II PROCESS MEASUREMENTS OPEN CHANNEL MONITOR Instruction Manual PL 269 February 1991 33452690 ...
Страница 6: ......
Страница 8: ......
Страница 12: ......
Страница 22: ......
Страница 28: ......
Страница 29: ......
Страница 30: ......
Страница 31: ......
Страница 32: ......
Страница 33: ......
Страница 34: ......
Страница 35: ......
Страница 36: ......
Страница 76: ......
Страница 78: ......
Страница 92: ...PL 269 9 14 ...
Страница 94: ......
Страница 95: ......
Страница 106: ......