When the tank is full, open the front cover to go into STANDBY
mode (I).
•
Let the system stand for 1 hour for effective bacteria elimination.
J
•
(After 1 hour) Check that the Tank Outlet Tubing is secured into the
drain.
Open the Tank Outlet Valve to drain all the water from the tank (J).
K
•
Close the Tank Outlet Valve (K).
L
•
Close the front cover (L).
The system will now go into FILLING TANK mode.
Let the tank level fill up to the 100% level.
•
Repeat steps I, J, K and L:
•
Open the front cover to go into STANDBY mode (I).
•
Check that the Tank Outlet Tubing is secured into the drain. Open
the Tank Outlet Valve to drain all the water from the tank (J).
•
Close the Tank Outlet Valve (K).
•
Close the front cover (L). The system will now go into FILLING
TANK mode.
The tank is now sanitized.