![Millipore direct-q 3 R Скачать руководство пользователя страница 53](http://html1.mh-extra.com/html/millipore/direct-q-3-r/direct-q-3-r_user-manual_1796915053.webp)
47
Flush Mode
ATTENTION
The Tank Outlet Valve should be left open during FLUSH mode.
H
•
The system will now go into FLUSH mode for 15 minutes (H). This is
done to empty the SmartPak cartridge of air and hydrate the
material inside.
I
•
When FLUSH mode is finished, the system will go into FILLING TANK
mode automatically (I).
Rinsing the SmartPak cartridge
J
•
Let FILLING TANK mode run for a minimum of 2 hours with the Tank
Outlet Valve open and placed to the drain. This will completely rinse
the purification media inside the SmartPak cartridge.
NOTE:
for sensitive applications, it is recommended to leave the system in
FILLING TANK overnight to ensure complete rinsing of the RO
membrane.
•
Close the Tank Outlet Valve (J) and connect it back to the system if
needed. The tank will start to fill up with water. It could take
approximately 1.5 to 2 hours to fill the tank. When it is full, the
Display will indicate a full tank by illuminating the symbols
representing the Tank Level
(see Section Understanding the Display)
•
The Vent Filter needs to be installed. See
Section Installation of the