59
How to Sanitize the Tank only
A tank sanitization is performed to eliminate bacteria growth in the tank only. It is recommended to sanitize the
tank at least once a year.
Things to Know BEFORE you sanitize the Tank
HAZARD
All safety precautions must be followed when handling Hydrogen Peroxide.
Rubber gloves, safety goggles and a lab coat must be worn to avoid any skin
and body contact.
A
•
If you have performed a system sanitization, then you do not need
to perform a tank sanitization.
•
You will need a minimum Tank Level display of at least 50% (A).
B
•
Locate the clear elbow fitting, the clear tubing and the syringe in
the Sanitization Kit (B).
•
You will need 200 ml of 30% Hydrogen Peroxide solution and
200 ml of purified water. Millipore SAS does not sell Hydrogen
Peroxide but it is readily available through most Scientific Supply
Companies.
•
The Vent Filter will need to be replaced after the sanitization is
completed.
•
The system will not be able to deliver Product Water while the
system is being sanitized.
ATTENTION
The total time needed to sanitize the tank is at least 4.5 hours.