How to Calibrate the Flowrate from the Direct-Q R (F02)
Before calibrating the Product flowrate from the Direct-Q R, you will need a 1 Litre graduated cylinder to measure
the total volume of water that will be dispensed. The Final Filter should be installed.
1
Press the Main and “-“ Buttons together
to enter the menu. The Display will show
F01.
2
Press the Main Button once. The Display
will show F02 and the 60 second timer.
The graduated cylinder will be blinking.
3
Press the Dispensing Button once. The
system will dispense water for 60 seconds.
The Display will show the timer counting
down.
NOTE:
If you press the Dispensing Button
again or if the Tank Level display is at
the 10% level before 60 seconds have
elapsed, then the system will stop
dispensing water. Calibration of the
flowrate is cancelled.
Continued on next page