Problem solving guide
101
Problem
Cause and remedy
Water in the pressure
chamber
Possible causes for water in the pressure chamber are as
follows:
- Condensate formation due to increased humidity in the
compressed air
- Defect in the on-site compressed air system, e.g. at the
water separator
- Condensation in the on-site compressed air line
- Defective seals on the insert
Check that your compressed air system is working cor-
rectly and for possible defects.
Have the insert checked and serviced by the Miele Cus-
tomer Service Department.
Next service due on:
This is not a fault.
The Miele Customer Service Department has recommen-
ded a date for the next service visit.
Please contact the Miele Customer Service Department
to arrange a service visit.
Dispensing/Dispensing systems
Caution when handling chemical agents.
For all chemical agents, the chemical agent manufacturer's safety instructions as given
on their safety data sheets must be observed.
Problem
Cause and remedy
The dispenser for powder
cleaning agent contains
residual agent at the end
of the programme.
The dispenser was still damp when cleaning agent was ad-
ded.
Make sure the dispenser is dry before adding powder
cleaning agent.
The dispenser flap was blocked by items in the cabinet.
Rearrange the load so that the flap can open.
The dispenser flap will not
close.
Residual cleaning agent is blocking the catch.
Remove the cleaning agent.
DOS Refill
During a programme sequence a low level of liquid chem-
ical agent in a container has been identified.
Replace the empty container with a full one.
Programme could not be
started. Vent DOS
A programme cannot be started because ...
... there is air in the dispensing system.
... the dispensing system has been sucked completely dry.
Check the level in the supply container. Replace an
empty container with a full one, if necessary.
Vent the dispensing system.
Dispensing system DOS
venting
This is not a fault.
The dispensing system will now be automatically vented.
Wait until the venting process is finished.