Chapter 7 Installation
Note:
Example to explain the relationship between the warmup time and the record interval (warmup time > record
interval):
Basic setting
Record interval
1min.
Warmup time
90 seconds
Output on the device
Sensor supply
Explanation: The sensor supply is activated when the time for the first recording following the PowerOn is reached.
As the warmup time is longer than the record interval, a measurement and recording cannot be completed at this
time. In the current example, a measurement and recording can only be completed after the second point in time
for the recording. As the sensor supply in this example is permanently active for the recording following the first
point in time, the measurement and recording are always completed at the exact recording time from the third
point in time.
Rev. 01
51