8.4
Test Block
A test block is included with the Portaflow equipment to allow the transducers and inter-connecting
cables to be functionally checked.
1. Switch ON the instrument.
2. Select
Quick Start
and enter the parameters shown in the table below for the appropriate
transducer type (A or B):
Parameter
A Sensors
B Sensors
Pipe outside diameter
30.0mm
50.0mm
Pipe wall thickness
14.0mm
22.0mm
Pipe lining thickness
0.00
Pipe wall material
Plastic
Fluid type
Water
Mode
Diagonal
Temp
20°C
3. At the end of the
Quick Start
procedure (see
page 19), the
Summary
screen is displayed.
Press the UP or DOWN arrow button. The
Sensors
screen is displayed.
4. Use the UP/DOWN scroll keys to select
Sensor Set
. Press the ENTER key.
5. Select the appropriate sensor (the default will
be "A") and press the ENTER key.
6. Select
Sensor mode
, choose
Diagonal
and
press the ENTER key.
7. Select
Return to Summary screen
and press
the ENTER key.
8. Check that the3 parameters displayed are correct.
9. Apply acoustic couplant to the sensors and attach them to the test block with the connectors
positioned towards the centre of the test block as shown in Figure x, and temporarily secure
them in place using elastic bands or tape.
10. Connect the sensors to the Portaflow using the cables provided (see page 3).
11. Press the ENTER key to display the
Flow Reading
screen.
12. Press the SYSTEM key (2) to display the
System Settings
screen.
13. Set
Damping
to at least 10 seconds.
14. Select
Save Setup & Exit
and press the ENTER key to return to the
Read Flow
screen.
15. The flow reading value displayed is not important. The fact that a reading is obtained indicates
that the instrument is functioning. This value may fluctuate but this is normal.