Response: Enter a valid number.
Wall thickness out of range
Interpretation: You have entered an out-of-range value for the pipe
wall thickness dimension
–
accepted range is 1mm - 75mm.
Response: Enter a valid number.
Lining thickness out of
range
Interpretation: You have entered an out-of-range value for the lining
thickness dimension
–
acceptable range is 0mm - 25mm.
Response: Enter a valid number.
Temperature range
Interpretation: You have entered an out-of-range value for the fluid
Temperature. Accepted temperature range -20°C to +300°C.
Response: Enter a valid number.
Invalid Date or Time
Interpretation: The entered Date or Time is invalid, or when setting up
‘timed’ data logging the Stop time is set earlier than the Start time.
Response: Enter a valid Date and Time.
Sensors: INVALID
Interpretation: The selected temperature is higher than the maximum
allowed for the sensor type.
Response: Select alternative sensors or change the temperature.
Mode: Err Typ
Interpretation: The selected sensors are invalid and the mode cannot
be verified.
Response: Select a valid sensor type and choose a mode that gives a
non-zero separation distance.