Micronics
Portaflow 222 & 333 User Manual
Issue 3.1 Feb 20
Page 15
2.4
Connecting Outputs
The output cable provides a 4-20mA current source that can drive a maximum load of
620Ω
and 3
digital output pairs for pulse or alarm outputs.
The isolated pulse output is provided by a SPNO/SPNC MOSFET relay which has a maximum load
current of 500mA and maximum load voltage of 24V AC/DC.
This output is suitable for SELV circuits only.
The pulse output is a Volt, or potential free contact and, when selected as a low flow alarm, is
configurable NO/NC.
The 4-20mA current output is available on the white (high polarity) and black (low polarity) wire pairs
(pins 1&2 on the connector). The alarm current due to a flow outside the range specified or due to a
loss of signal is set at 2.5mA.
Using the instrument’s menu system (see page
34), you can:
Select the current output function Off/On
Select the current output signal range (4-20mA, 0-20mA, 0-16mA)
Calibrate the current output signal to a required flow range
Figure 13
Output Wiring
The three digital output pairs for pulse or alarm outputs are as follows:
Output
Pin outs
Wire pair
Output 1
Pins 3 & 4
Brown/White
Output 2
Pins 5 & 6
Red/White
Output 3
Pins 7 & 8
Orange/White