6.1.1
Converting the measured current to flow rate
Assume the maximum flow rate is F
max
(l/min) and the minimum flow rate F
min
is ‘0’ (l/min), as shown
below.
To calculate the flow rate (l/min) for a measured current (mA) then:
0-20mA
0-16mA
4-20mA
𝐹𝑙𝑜𝑤 𝑟𝑎𝑡𝑒 = 𝐼 𝑥
(𝐹
𝑚𝑎𝑥
− 𝐹
𝑚𝑖𝑛
)
20
+
𝐹
min
𝐹𝑙𝑜𝑤 𝑟𝑎𝑡𝑒 = 𝐼 𝑥
(𝐹
𝑚𝑎𝑥
− 𝐹
𝑚𝑖𝑛
)
16
+
𝐹
min
𝐹𝑙𝑜𝑤 𝑟𝑎𝑡𝑒 = (𝐼 − 4) 𝑥
(𝐹
𝑚𝑎𝑥
− 𝐹
𝑚𝑖𝑛
)
16
+
𝐹
min
6.2
Digital Outputs
The three digital outputs can each be set up to operate in one of three modes:
Pulse Output (set to
Normally Open
or
Normally Closed
contact types
)
Alarm Output (set to
Rising
or
Falling
signal directions)
Frequency Output (with
High Frequency
and
Low Frequency
settings)
The measurement source can be:
Volume (not compatible with Frequency Output)
Flow (not compatible with Pulse Output)
Energy (not compatible with Frequency Output)
Power (not compatible with Pulse Output)
Signal (not compatible with Pulse Output)
To configure any of the digital outputs:
1. From the MAIN menu, use the Up and Down scroll keys to select
Setup Instrument
. Press the
ENTER Key. With
Output
selected in the
Options
menu, press the ENTER Key.
Alternatively, from a
Read Flow/Velocity/Energy
screen, press the OUTPUTS key (3). The
Output Board
menu is displayed.