8. Stand the sample tubes on the sample tube rack and place the filler rods in a basket or in
the rack. Bake in a vacuum oven for two hours at 110 °C.
Samples tubes can also be cleaned with high purity acetone or isopropyl alcohol and
dried for about 10 minutes under heat. If using this method, continue with step 10.
9. Remove the sample tubes and filler rods from the oven and allow to cool.
Do not insert the filler rods at this time. Filler rods are inserted before the sample tube
is installed on the analysis port.
10. Blow out the sample tubes with oil-free compressed air.
11. Rinse the sample tube closure with isopropyl alcohol, then wipe the sample tube closure dry
with a clean, lint-free cloth.
12. Label the sample tube and stopper for identification.
13. Replace the rubber stopper.
Clean and Label Sample Tubes
Gemini VII Keypad Operator Manual
239-42801-01 (Rev E) — July 2021
2 - 7