5 Troubleshooting
R
EPLACE A
G
AS
C
YLINDER
A. Gas tubing to instrument
B. Gas regulator shut-off valve
C. Low pressure gauge
D. High pressure gauge
E. Gas cylinder shut-off valve
F. Regulator connecter nut
G. Regulator control knob
H. Brass reducer fitting
Disconnect a Depleted Gas Cylinder
1. Close the regulator shut-off valve and gas cylinder shut-off valve by turning the knobs clock-
wise.
2. Disconnect the gas line from the regulator. Gas will be vented from the line. It is not neces-
sary to disconnect the gas line from the analyzer inlet if the cylinder will be replaced imme-
diately with one of the same type.
3. Open the gas regulator shut-off valve by turning the knob counterclockwise. Gas will be ven-
ted from the regulator.
4. Turn the regulator control knob clockwise to open and vent any remaining gas. Both gauges
should read at or near zero. If not, make sure the gas regulator shut-off valve is open.
5. Close the regulator by turning the control knob counter-clockwise.
6. Use an appropriate wrench to loosen the nut at the regulator connector nut then remove the
regulator from the cylinder.
7. Replace the protective cap on the depleted cylinder. Disconnect the retaining strap and
move the cylinder to an appropriate location.
Connect a Gas Cylinder
Analyzer
Gauge should indicate
Gemini
15-18 psig (103 - 124 kPag)
Regulator Pressure Settings
Exceeding the maximum recommended air pressure could cause personal injury or
damage the instrument.
5 - 8
Gemini VII Keypad Operator Manual
239-42801-01 (Rev E) — July 2021
Содержание GEMINI VII KEYPAD
Страница 1: ...OPERATOR MANUAL 239 42801 01 July 2021 Rev E GEMINI VII KEYPAD SURFACE AREA ANALYZER...
Страница 10: ......
Страница 14: ......
Страница 30: ......
Страница 52: ......
Страница 108: ......
Страница 146: ......
Страница 148: ......
Страница 154: ......
Страница 155: ......
Страница 156: ......