5 Troubleshooting
11. Press
ENTER
to start the analysis.
12. After the analysis is complete, the window shows the results. Compare the BET results in
the window with those shown on the label of the reference material bottle.
n
If the analysis results match those shown within the tolerance level on the bottle, the
analyzer is ready to use.
n
If the analysis results are not within the tolerance level, repeat steps 2 through 10 to
analyze the sample again. Verify that the value entered for the mass is correct. If the
analysis results still fail to match, contact a Micromeritics Service representative.
5 - 10
Gemini VII Keypad Operator Manual
239-42801-01 (Rev E) — July 2021