3 Command Prompts
Option
Description
Each prompt has a Yes/No response. If
Yes
is selected at any of these
prompts, enter the appropriate ranges.
After specifying Pass/Fail criteria, the prompts that display for the
Yes
choice is shown. If
Yes
is selected for the t-method report:
Enter the relative pressure at which to
start
selecting collected data
points for calculating t-method micropore volume.
Range:
0.0
to
1.0 P/Po
Default:
0.0 P/Po
Enter the relative pressure at which to
stop
selecting collected data
points for calculating t-method micropore volume.
Range:
t-method range from value to 1.0 P/Po
Default:
0.7 or t-method range from P/Po
If the range specified includes more than 50 data points, only the first 50
points are used in t-method calculations. Individual points after analysis
may be selected or deselected.
Report Options (continued)
3 - 24
Gemini VII Keypad Operator Manual
239-42801-01 (Rev E) — July 2021