4. Remove the O-ring from the top of the sample tube and replace it with a new one.
If the O-ring remains inside the sample port, use the O-ring removal tool from the
accessory kit
[tool part number 004-54618-00
].
5. After the new O-ring is in place, insert the sample tube back into the sample port until it is
fully seated.
6. Slide the sample tube connector nut up the tube until it comes in contact with the port fitting
(the ferrule and O-ring will move along with the connector nut). Then, turning clockwise,
hand tighten the connector nut to the sample connector.
Check and Clean the Dewar
Gemini VII Keypad Operator Manual
239-42801-01 (Rev E) — July 2021
5 - 15