3 Command Prompts
Field or Button
Description
Equilibrate
Selects the analysis mode.
Choices:
Equilibrate, Scan
Equilibration time
Increases the pressure in steps to the next specified level for data
collection. It maintains this pressure and monitors the volume
adsorbed. When the rate of adsorption (volume adsorbed during the
equilibration time divided by total volume adsorbed) falls below
0.01% and the variance from 11 consecutive readings falls below
0.1%, equilibration is assumed to have occurred. The analyzer then
introduces the next adsorbate gas dose.
Enter the equilibration time.
Range:
1
to
1000 sec
Analysis Condition Prompts (continued)
3 - 18
Gemini VII Keypad Operator Manual
239-42801-01 (Rev E) — July 2021