Error Codes
scanCONTROL 29xx
Flashing
sequence
Cause
Remedy
Notes
--..
2x long, 3x short
Data overflow during
transmission of the data via
Ethernet
Reduce profile frequency,
increase packet size.
Data can be faulty.
--…..
2x long, 5x short
Error during calculation
Reduce profile frequency, se-
lect faster calculation mode.
Data can be faulty.
--......
2x long, 6x short
Error during Ethernet trans-
mission
Reduce profile frequency.
Data can be faulty.
Group: Output Unit
---
3x long
scanCONTROL Output Unit
not found.
Connect the sensor with the
scanCONTROL Output Unit.
---
---.
3x long, 1x short
Connected modules of
scanCONTROL Output Unit
are not supported.
Use the supported modules
only, see Chap.
---
---..
3x long, 2x short
Communication error (scan-
CONTROL Output Unit)
Reduce profile frequency.
---
---...
3x long, 3x short
Output overflow (scanCON-
TROL Output Unit)
Reduce profile frequency.
---
Group: Ethernet interface
----
4x long
IP address conflict
Check the Ethernet configura-
tion of sensor and the host
PC.
Choose another IP address for
the sensor.
If the problem persists,
please contact the manu-
facturer.
The
state
LED flashes green; long during active data transmission and short for controller accesses. A
controller access can cause various data overflows particularly when the measuring frequency is near its
maximum.