Page 68
Appendix| EtherCAT Documentation
combiSENSOR 6430
Object 2200h: Commands
2200
RECORD
Commands
ro
Subindices
0
VAR
Number of entries
2
Unsigned8
ro
1
VAR
Command
AVT1
Visible String
rw
2
VAR
Command Response
AVT1OK
Visible String
ro
Any commands can be sent to the controller with the object 2200h, for example, the thickness functions as
these are not defined in the COE objects.
Object 6020h: Measuring values
6020
RECORD
Measuring values
ro
Subindices
0
VAR
Number of entries
11
Unsigned8
ro
3
VAR
Counter
xxxx
Unsigned32
ro
8
VAR
Channel 1
xxxx
Float
ro
9
VAR
Channel 2
xxxx
Float
ro
10
VAR
Channel 3
xxxx
Float
ro
11
VAR
Channel 4
xxxx
Float
ro
A 2.3
Measurement Data Format
The measuring values are transmitted as Float.
The unit can be read from the channel info objects 2020h to 2023h (unit).