Page 51
Ethernet Interface
combiSENSOR 64x0
Thickness
measurement
Dielectric constant ε
r
Value
Dielectric constant of the target used. When the di-
electric constant is calculated, the controller replaces
the original value.
Offset
Value
Standard value: 0. Adds a constant [µm] to the thick-
ness value.
Max. working distance
sensor
Value
Working distance sensor [µm], value range: 5000 µm
or 10000 µm.
Clear
This button deletes the thickness measurement func-
tion, channel 1 outputs the difference = Eddy - Capa.
Zero thickness
This button sets the value of the offset in such a way
that the thickness signal is 0 µm.
Submit
Clicking onto this button permanently stores the
values in the controller. Subsequently, the controller
calculates based on the new values.
Calculate dielec-
tric constant ε
r
Difference signal without
target
Value
Step 1: Remove the measurement object from the
measuring gap, click onto the button Set current
position.
Difference signal with
target
Value
Step 2: Move the measurement object into the
measuring gap, click onto the button Set current
position.
Max. working distance
sensor
Value
Step 3: Enter the value [µm] for the maximum working
distance. Value range: 5000 µm or 10000 µm.
Thickness object
Value
Step 4: Enter the thickness value [µm] of the master
target.
Dielectric constant ε
r
Value Calculated dielectric constant
Calculate
Clicking onto the button starts the calculation of the
dielectric constant
ε
r
.
Содержание combiSENSOR 64x0
Страница 1: ...Instruction Manual combiSENSOR 64x0...
Страница 6: ...combiSENSOR 64x0...
Страница 73: ......