Page 46
Installation and Initial Operation
thicknessCONTROL MTS 8202.LLT / 4350127.347
5.11.2 Check Measuring Device Capability, Method 1
Start measurement.
Change to menu
Extras > CGM Method 1
.
Based on a defined point in the cross-profile, 50 measurements are analyzed.
Fig. 58 Measurement of measuring device capability, Method 1
1
Result of the measuring device capability - valid after 50 points: Cgm / Cgk > 1.33
T(min): calculated tolerance
2
This diagram shows the measurement values. After each completed scan, a
measuring point is added.
3
Setpoint, upper and lower limit of production. Setpoint and threshold value are
used for calculation and diagrams.
4
Value to be checked [x]: This option defines the measuring point to be checked
[1 .. up to x] – this point is also shown as cursor.
5
This button starts and ends the test.
6
Shows the current state:
Test ongoing
Test completed, result not OK
Test completed , result OK
7
The result of the test is saved in an Excel file.
8
This diagram shows the current cross-profile.
Содержание 4350127.347
Страница 1: ...thicknessCONTROL MTS 8202 LLT C frame Non contact thickness measurement Operating Instructions...
Страница 10: ...Page 10 Overview Technical Data Safety Instructions thicknessCONTROL MTS 8202 LLT 4350127 347...
Страница 35: ...Page 35 Installation and Initial Operation thicknessCONTROL MTS 8202 LLT 4350127 347 Fig 39 System parameter...
Страница 56: ...Page 56 Installation and Initial Operation thicknessCONTROL MTS 8202 LLT 4350127 347...
Страница 64: ...Page 64 Maintenance Service Spare Parts Lists thicknessCONTROL MTS 8202 LLT 4350127 347...
Страница 66: ...Page 66 Declaration of Incorporation thicknessCONTROL MTS 8202 LLT 4350127 347...