Veri-Q PREP M16 G2-16TU
User Manual
12. Cleaning
12.1 Cleaning of Air Rubber
12.1.1
Open the upper tray.
12.1.2
Prepare a clean cotton or cloth wetted with 70% ethanol.
12.1.3
Scrub the air rubber, especially the bottom side, with cotton or cloth wetted with 70%
ethanol.
12.2 Cleaning of Tray
12.2.1
Open the upper tray.
12.2.2
Prepare a clean cotton or cloth wetted with 70% ethanol.
12.2.3
Wipe the top and bottom side of the upper tray. And if necessary, wipe the hole of the
upper tray with a cotton swab.
12.2.4
Open even the lower tray and wipe it in the same manner.
12.3 Cleaning of Pre-run Cartridge
12.3.1
Dispose the waste liquid from the ‘pre-run cartridge.’
12.3.2
Wipe it clean with cotton or cloth wetted with 70% ethanol.
12.3.3
Dry the cartridge at room temperature.
12.4 Cleaning of Solution Bottle
12.4.1
Check solution bottle on whether there is particle or dust inside.
12.4.2
If there is, dispose the solution and clean the bottle with distilled water.
12.4.3
Dry the bottle under the clean bench.