Veri-Q PREP M16 G2-16TU
User Manual
4.3 Transporting the G2-16TU
When moving the device, please protect the device by using the silver case. When packing the
equipment,
•
Remove all liquid in a bottle
•
Perform Pre-run 3 times to remove liquid in the equipment
•
Place the power cable and AC adapter at the bottom of the case
•
Put the equipment in the silver case
•
Put the pre-run cartridge in the empty space on the side of the case.
4.4 How to Use
Check the G2-16TU device with the required parts according to installation protocol. You must
start the operation in a normal operating condition.
4.1.1
Power Supply and Self-Test Method
1) Connect instrument power cable on the power connect at the rear side of the G2-16TU, and
then, turn on the power.
2) When power is supplied to the device, POWER LED should light up. Please refer to the
following photo. The green light coming on means the device is working properly.
3) When the power is turned on, self-test and initialization is automatically performed. If a
problem occurs during a self-test, the equipment does not work properly, and a message,
“Motor Initialize (RUN-INIT)” appears on the bottom of the LCD. In this case, do not operate
the equipment by force. And please make an inquiry to the AS team of Mico Biomed Co., Ltd.
regarding such malfunction.
4.5 G2-16TU Nucleic Acid Extraction Procedure
For sample preparation, see the instructions for use of Veri-Q PREP M16 reagent kit made by
MicoBioMed
4.5.1
Filling buffer solution
1) Carefully fill up each bottle with the solution. Afterwards, place the bottle in a suitable location.
2) Open the lid after suitably placing the bottle.