Veri-Q PREP M16 G2-16TU  

User Manual

4.3  Transporting the G2-16TU

When moving the device, please protect the device by using the silver case. When packing the 
equipment,

Remove all liquid in a bottle

Perform Pre-run 3 times to remove liquid in the equipment

Place the power cable and AC adapter at the bottom of the case

Put the equipment in the silver case

Put the pre-run cartridge in the empty space on the side of the case.

4.4  How to Use 

Check the G2-16TU device with the required parts according to installation protocol. You must 
start the operation in a normal operating condition. 

4.1.1  

Power Supply and Self-Test Method

   1)  Connect instrument power cable on the power connect at the rear side of the G2-16TU, and 

then, turn on the power.

   2)  When power is supplied to the device, POWER LED should light up. Please refer to the 

following photo. The green light coming on means the device is working properly.

   3)  When the power is turned on, self-test and initialization is automatically performed. If a 

problem occurs during a self-test, the equipment does not work properly, and a message, 
“Motor Initialize (RUN-INIT)” appears on the bottom of the LCD. In this case, do not operate 
the equipment by force. And please make an inquiry to the AS team of Mico Biomed Co., Ltd. 
regarding such malfunction.

4.5  G2-16TU Nucleic Acid Extraction Procedure 

For sample preparation, see the instructions for use of Veri-Q PREP M16 reagent kit made by 
MicoBioMed

4.5.1  

 Filling buffer solution

   1) Carefully fill up each bottle with the solution. Afterwards, place the bottle in a suitable location.
   2) Open the lid after suitably placing the bottle.

Содержание 9S101

Страница 1: ...Veri Q PREP M16 G2 16TU User Manual Apr 2020 Cat No 9S101...

Страница 2: ...16TU For safe use please use the device after reading the manual thoroughly Please furnish this document at a designated location so it can be used at all times MiCo BioMed Co Ltd 3rd and 4th Floor 54...

Страница 3: ...3 How to Obtain More Information 6 1 4 About Customer Support 6 1 5 Precautions on distribution or duplication of the manual 6 1 6 Laboratory Biosafety 6 2 Safety Guideline 7 2 1 Precautions when usi...

Страница 4: ...pleting a training course Servicing on G2 16TU is performed by MiCo BioMed 2 1 3 The user must use only components of MiCo BioMed in order to receive guarantee of the product Please perform regular ma...

Страница 5: ...ppropriate data values 2 1 5 If the device does not meet the performance conditions make an inquiry to the manufacturer 2 1 6 This device is an electronic and mechanical device Since there is a risk o...

Страница 6: ...or fumes and disposal of wastes must follow all national and local health and safety regulations and laws Please wash your hands after handling the reagents 2 2 19 Operate the product in an environmen...

Страница 7: ...each parts Near sample tubes Biological hazard Since the tubes or waste cartridges become contaminated with biohazardous materials they must be handled while wearing gloves Near buffer bottles and sam...

Страница 8: ...tantaneous maximum power consumption 110W Safety device We made sure the electric shock inside the board does notaffect the element by using Varistor on operating parts such as valve pump fan etc We m...

Страница 9: ...l be filled in this waste cartridge Lid This lid will cover the Elution Cartridge 3 6 Equipment Part Names and Functions 1 2 3 4 5 6 7 No Name Explanation 1 LCD Operates and sets equipment with touch...

Страница 10: ...y Fluid control part for washing air pumping and extraction Moving part to perform pre treatment per channel 4 1 2 Operation Principle This device has the function to rapidly extract nucleic acid from...

Страница 11: ...e G2 16TU and then turn on the power 2 When power is supplied to the device POWER LED should light up Please refer to the following photo The green light coming on means the device is working properly...

Страница 12: ...place a Pre run Cartridge in an appropriate position of Sample Rotor Do not place the Pre run cartridge upside down The buffer solution can flow into the equipment Please refer to the photo below X O...

Страница 13: ...nge menu page by pressing image or pressing button You can check the firmware version on the left corner The latest version is v2 00 4 5 8 Select extraction volume 1 Press E Volume image or button You...

Страница 14: ...ration 4 5 9 Start operation 1 Press START button or Start image The equipment automatically operates Please refer to the following for operating procedures A Checking Binding When the user operates t...

Страница 15: ...e visit http micobiomed com or send an e mail to salesbio micobiomed com C Elution In this step the eluate will be gathered in the micro tube which is inserted into the elution cartridge Because G2 16...

Страница 16: ...ipment b When sample extraction is complete In this case take the following measures a Remove all sample tubes from the upper tray b Press INIT image or SET button Washing1 Buffer Empty This error mes...

Страница 17: ...ed purchase information please check the homepage http micobiomed com 7 Warranty Policy MiCoBioMed shall be excluded from warranty obligations when a person other than the dedicated engineer performs...

Страница 18: ...of the upper tray And if necessary wipe the hole of the upper tray with a cotton swab 12 2 4 Open even the lower tray and wipe it in the same manner 12 3 Cleaning of Pre run Cartridge 12 3 1 Dispose...

Страница 19: ...ound the equipment The device is not exposed to direct sunlight There are no installations that can cause dust heat and fluid inflow into the equipment The port switch direction on the rear of the equ...

Страница 20: ...not be required to replace any unit which is damaged or malfunctions due to abuse accidents alteration neglect misuse maintenance by someone other than Mico BioMed co Ltd or failure to operate in acco...

Страница 21: ...Veri Q PREP M16 G2 16TU MiCo BioMed Co Ltd 3rd and 4th Floor 54 Changeop ro Sujeong gu Seongnam si Gyeonggi do Korea 13449 TEL 82 70 5227 6000 FAX 82 70 5227 6001 2 User Manual Apr 2020...

Отзывы: