Veri-Q PREP M16 G2-16TU
User Manual
4.5.3
Placing Elution Tube and Cartridge
1) Open the lower tray and place a Waste cartridge.
2) Insert the Elution Cartridge: Place the Elution Cartridge so it fits in the location for insertion.
3) Insert Elution tube: Insert the Elution tubes into the Elution Cartridge.
Make sure the lid of each tubes faces outward.
4.5.2
Pre-running
1) Please check “PreRUN Required” message from the configuration screen.
2) Pull out the upper tray and place a “Pre-run Cartridge” in an appropriate position of “Sample
Rotor.”
** Do not place the Pre-run cartridge upside down.
The buffer solution can flow into the
equipment. Please refer to the photo below.
X
O
Correct
Incorrect
A. Press “Prerun” button on the LCD screen.
B. If “YES” button is pressed, “Prerun” begins (It will take about ~80 seconds.)
C. If you hear a beep sound, the equipment automatically finishes “Prerun.”
D. Press “YES” button and then, remove “Pre-run Cartridge.”
E. Remove buffer solution from the waste liquid container.
F. Refill the bottle with solution
**Please fill the bottle with solution after Pre-run is finished.
**Please perform pre-run right before starting sample extraction.