XTP601 User’s Manual
10
97313 Issue 5, May 2018
OPERATION
2.5.1
Main Page
20.91
% 02
Cell T Not Stable
Back. Gas: N2
Figure 10
Main Page
Parameter
Description
Background Gas
Displays the background gas that the unit was calibrated in
% O2
(if HSR=OFF)
Real oxygen reading in %
Display resolution = 0.01
(Display resolution 0.1% when min value is >=20%)
% O2p
(if HSR = ON)
HSR oxygen reading (extrapolated quick response value of real
oxygen reading)
Real oxygen value is displayed in status bar
Flashing “Heating”
Symbol
This flashes until cell temperature is stable at set point ±0.5°C for
minimum of 5 minutes
Status Bar
Flashes any system warnings and error messages (see below)
Status Message Table
Message
(Trigger Condition)
Light Guide
(if fitted)
% O2 out of range
(beyond calibration range, e.g. 0-25%)
N/A
AL1 ON
ORANGE1 ON (app s/w only)
AL2 ON
ORANGE2 ON (app s/w only)
Comp i/p signal error
(input < 3.6 mA or > 21 mA)
RED FLASH
(priority2)
Ext sens signal error
(input < 3.6 mA or > 21 mA)
RED FLASH
(priority2)
Cell T not stable
(not within ±0.5ºC of set point for continuous period
of 5 minutes)
RED ON
(priority1)
Cell T sensor error
(cell temp measures <-50 or >+80ºC)
RED ON
(priority1)
Press sensor error
(pressure sensor < 850 or >1100 mbar)
RED ON
(priority1)
O2
sensor error
(Vcomp <1 or >8191)
RED ON
(priority1)
PCB temp too high
(PCB temp > cell temp set point)
RED ON
(priority1)
Содержание XTP601
Страница 43: ...XTP601 User s Manual Michell Instruments 35 APPENDIX A Appendix A Technical Specifications...
Страница 47: ...XTP601 User s Manual Michell Instruments 39 APPENDIX B Appendix B Modbus Register Map...
Страница 54: ...XTP601 User s Manual 46 97313 Issue 5 May 2018 APPENDIX C Appendix C Hazardous Area Certification...
Страница 61: ...XTP601 User s Manual Michell Instruments 53 NOTES...
Страница 62: ......
Страница 63: ......
Страница 64: ...http www michell com...