XTP601 User’s Manual
16
97313 Issue 5, May 2018
OPERATION
2.7.2 Human Machine Interface (HMI) Page
CONTRAST
O-100%
O-100%
C/F/K
psia, bara, kpa
2-60s
DD/MM/YY
BRIGHTNESS
TEMPR UNIT
EXT PRESS UNIT
CHART INTVAL
DATE
Figure 18
HMI Page
It is possible to change parameters within the HMI, as shown below:
Setting
Description/Operation
Options
CONTRAST
LCD contrast setting
0-100%
in 10% steps
BRIGHTNESS
LCD backlit setting
0-100%
in 10% steps
TEMPR UNIT
Global temperature unit selection
ºC, ºF, K
EXT PRESS UNIT
Pressure unit selection (for external
sensor only)
psia, bara, kPa
CHART INTVAL
Chart interval
2-60 s
in 2 sec steps
DATE
Date on the LCD can be either
format
DD/MM/YY or
MM/DD/YY
Scroll down past the DATE field to access the clock page.
2.7.3 Clock
Page
HOURS
00-23
00-59
1-31
1-12
00-99
**:**:**
MINS
DAY
MONTH
YEAR
LIVE CLOCK
Figure 19
Clock Page
The real time clock and calendar is used to store date/time information for log data,
min/max data and date of calibration. On entering this page all fields are initialized with
the current values. These can be also be set through the application software.
Setting
Description/Operation
Options
HOURS
Hours
00-23
MINS
Minutes
00-59
DAY
Day
1-31
MONTH
Month
1-12
YEAR
Year
00-99
LIVE CLOCK
Current Time
**:**:**
Содержание XTP601
Страница 43: ...XTP601 User s Manual Michell Instruments 35 APPENDIX A Appendix A Technical Specifications...
Страница 47: ...XTP601 User s Manual Michell Instruments 39 APPENDIX B Appendix B Modbus Register Map...
Страница 54: ...XTP601 User s Manual 46 97313 Issue 5 May 2018 APPENDIX C Appendix C Hazardous Area Certification...
Страница 61: ...XTP601 User s Manual Michell Instruments 53 NOTES...
Страница 62: ......
Страница 63: ......
Страница 64: ...http www michell com...