XTP601 User’s Manual
Michell Instruments
11
OPERATION
2.5.2 Chart
Page
t
t-4
25
%
0
Figure 11
Chart Page
NOTE: This data is not available via the Modbus
•
This indicative chart is continuously running at the set chart interval (2
to 60 seconds).
•
Chart duration in seconds = (chart interval * 60).
•
It is auto ranging with an auto range resolution of 1% O
2
.
•
It is reset if the chart interval is changed, the Display is changed between
% O
2
and
% O
2
HSR
or the instrument power is cycled.
•
The chart will record either % O
2
(Non HSR) or % O
2
HSR if this is set to
ON
in the setup pages.
•
Chart buffer size is 60 values and therefore the chart duration changes
according to interval chosen.
•
Chart data is only stored in volatile memory and therefore is not saved.
Chart interval is saved and available in a Modbus register.
•
Chart data is not available via serial comms as the Application Software is
able to perform more sophisticated charting functions.
Содержание XTP601
Страница 43: ...XTP601 User s Manual Michell Instruments 35 APPENDIX A Appendix A Technical Specifications...
Страница 47: ...XTP601 User s Manual Michell Instruments 39 APPENDIX B Appendix B Modbus Register Map...
Страница 54: ...XTP601 User s Manual 46 97313 Issue 5 May 2018 APPENDIX C Appendix C Hazardous Area Certification...
Страница 61: ...XTP601 User s Manual Michell Instruments 53 NOTES...
Страница 62: ......
Страница 63: ......
Страница 64: ...http www michell com...