![Michell Instruments XTP601 Скачать руководство пользователя страница 29](http://html1.mh-extra.com/html/michell-instruments/xtp601/xtp601_user-manual_1784648029.webp)
XTP601 User’s Manual
Michell Instruments
21
OPERATION
2.7.9 Light Guide Option
The light guide is an optional accessory fitted, ideally, to the left hand cable entry and
has a red and green LED to display the status.
•
Green On - indicates instrument power is on.
•
Red LED flashing - indicates when external compensation input or
external sensor is out of range (if either is selected to
ON
). Out of range
is <3.6 mA or >21 mA (see Status Message Table in Section 2.5.1).
•
Red LED On – indicates an internal sensor error or instrument cell
temperature not yet stabilized (see Status Message Table in Section 2.5.1)
– for LED indication this takes priority over external errors.
These have been designed to follow the NAMUR NE44 standard.