XTP601 User’s Manual
32
97313 Issue 5, May 2018
INSTALLATION
4.5.3 Signal
Output
There are two 4-20mA linear signal output channels. Both are for oxygen concentration.
One is fixed on the calibrated range of the unit and the second can be configured
in the menu.
NOTE: When the instrument is warming up (cell temperature
not stabilized) these outputs are factory set to 3.2 mA to indicate that the
instrument is not ready.
•
The maximum mA output is approximately 20.5mA
•
The minimum mA output is approximately 3.8mA
•
The user can select the falut conditon to drive the mA output Low (3.2mA)
or Hi (21.4mA).
4.5.4 Serial
Output
The analyzer has Modbus RTU communications over RS485 protocol; please see
Application Software CD for more details.
•
Type: Modbus RTU over RS485
•
RS485: 2 wire (plus ground), half duplex
•
Baud Rate: 9600
• Parity:
None
•
Data bits: 8
•
Stop bits: 1
4.5.5 Analog (4-20 mA) Outputs and Communications (PL5 - Green)
Warning: Do not attempt to loop-power this instrument via the 4-20 mA
output as this will irreversibly damage the main pcb.
PIN 7
PIN 6
PIN 5
PIN 4
PIN 3
PIN 2
PIN 1
RS485
GND
RS485
B
RS485
A
Ch2
O/P
-
Ch2
O/P
+
Ch1
O/P
-
Ch1
O/P
+
NOTE 1: Channel 1 is fixed range output over instrument range and Channel
2 is adjustable between 0-100%.
NOTE 2: For long cable lengths, a termination resistor of 120
Ω
is provided to
aid in impedance matching of the communications cable. This is selectable
via JMP3. A terminating resistor is simply a resistor placed at the extreme
end or ends of a cable on an RS485 network to reduce effects of impedance
mismatch. Impedance mismatch can cause reflections of data as it passes
down the cable and the reflections could be large enough to cause data errors.
NOTE 3: The maximum loop load resistance for mA outputs is 550
Ω
NOTE 4: the mA outputs are powered with a 24V excitation voltage.
Содержание XTP601
Страница 43: ...XTP601 User s Manual Michell Instruments 35 APPENDIX A Appendix A Technical Specifications...
Страница 47: ...XTP601 User s Manual Michell Instruments 39 APPENDIX B Appendix B Modbus Register Map...
Страница 54: ...XTP601 User s Manual 46 97313 Issue 5 May 2018 APPENDIX C Appendix C Hazardous Area Certification...
Страница 61: ...XTP601 User s Manual Michell Instruments 53 NOTES...
Страница 62: ......
Страница 63: ......
Страница 64: ...http www michell com...