![Michell Instruments XTP601 Скачать руководство пользователя страница 55](http://html1.mh-extra.com/html/michell-instruments/xtp601/xtp601_user-manual_1784648055.webp)
XTP601 User’s Manual
Michell Instruments
47
APPENDIX C
Appendix C
Hazardous Area Certi
fi
cation
The XTP601-EX Oxygen Analyzer is certified compliant to the ATEX Directive (2014/34/
EU) and IECEx for use within Zone 1 & 2 Hazardous Areas and has been assessed so by
SIRA Certification (Notified Body 0518).
The XTP601-EX Oxygen Analyzer is certified compliant to the North American Standards
(USA and Canada) for use within Class I, Division 1, Groups B, C and D Hazardous
Locations and has been assessed so by cCSAus.
C.1 Product
Standards
This product conforms to the Standards:
EN60079-0:2009 IEC60079-0:2007
EN60079-1:2007 IEC60079-1:2007
EN60079-31:2009 IEC
60079-31:2008
CSA C22.2 No 0-10
UL 916
CSA C22.2 No 30-M1986 (R 2007)
UL 1203
CSA C22.2 No 142-M1987 (R 2009)
C.2 Product
Certi
fi
cation
This product is attributed with the product certification codes:
ATEX & IECEx
II 2 GD Ex d IIB+H2 T6 Gb
Ex tb IIIC T85°C Db IP66
North American
Class I, Division 1, Groups B, C, D
C.3 Global
Certi
fi
cates/Approvals
ATEX
Sira12ATEX1042X
IECEx
SIR12.0013X
cCSAus
2541173
TC TR Ex
RU C-GB.
ГБ
05.B.00152
INMETRO
NCC 14.03166 X
These certificates can be viewed or downloaded from our website at:
http://www.michell.com
C.4 Special
Conditions
1.
Cable entry holes shall be fitted with either an appropriately certified
cable gland or appropriate certified blanking element. These shall
provide and maintain a minimum enclosure ingress protection of IP66.
2. The maximum pressure associated with the process medium in the
internal pipes shall be limited to 4 bar.
3.
The maximum temperature associated with the process medium shall
be limited to 55°C.
4.
The flameproof joints of the flame arrestors & breather are not intended
to be repaired.
Содержание XTP601
Страница 43: ...XTP601 User s Manual Michell Instruments 35 APPENDIX A Appendix A Technical Specifications...
Страница 47: ...XTP601 User s Manual Michell Instruments 39 APPENDIX B Appendix B Modbus Register Map...
Страница 54: ...XTP601 User s Manual 46 97313 Issue 5 May 2018 APPENDIX C Appendix C Hazardous Area Certification...
Страница 61: ...XTP601 User s Manual Michell Instruments 53 NOTES...
Страница 62: ......
Страница 63: ......
Страница 64: ...http www michell com...